Uncategorized

Performance of intracortical microelectrode arrays in people with implanted brain–computer interfaces over a 20-year period

Performance of intracortical microelectrode arrays in people with implanted brain–computer interfaces over a 20-year period

Between 2004 and 2019, 14 participants were enrolled in the BrainGate and BrainGate2 pilot clinical trials; each participant had one or two 96-channel platinum-metalized Utah microelectrode arrays placed in cortex. Overall, 19 of the total 20 arrays were placed in the hand area of the motor cortex (Fig. 1a,b), with one array placed in the middle frontal gyrus. The BrainGate and BrainGate2 clinical trials were designed to evaluate the safety of implanted microelectrode arrays and to discover fundamental insights into the effective design of intracortical BCIs. Testing new neural decoding architectures and task designs was common throughout the trials, and as such it was not feasible to summarize across all types of data collected.

Fig. 1: Datasets curated for assessing recording and decoding performance.
Fig. 1: Datasets curated for assessing recording and decoding performance.

a, One or two 96-channel intracortical microelectrode arrays were placed in the cortex of 14 participants as part of the BrainGate and BrainGate2 clinical trials. b, In total, 19 of 20 arrays were placed in the hand knob area of the dominant precentral gyrus (PCG), shown in red, and one was placed in the middle frontal gyrus (MFG) of participant T10. c, Example unsorted threshold-crossing waveforms, detected during a representative 10 second recording window. For this study, 30 kHz recordings were decimated to 15 kHz, band pass filtered with a pass band of 250–5,000 Hz, re-referenced with linear regression referencing and thresholded at −4.5-times the robust standard deviation of the voltage signal for each channel. For each threshold crossing, a 2.13 ms voltage snippet was extracted. d, To assess decoding performance, we retrospectively analyzed sessions in which participants were asked to attempt to move a computer cursor toward a defined target. Imagery varied between sessions and participants. Tasks included Fitts, Radial-8, grid tasks and their variations. To assess array recording yields, data from all types of sessions were used (including sessions where no cursor control occurred). e, Example cursor trajectories for a radial-8 task. f, An overview of recording durations and sessions analyzed. Total array implant days ranged from 296 to 2,780, with a median of 774 days (n = 20 arrays). After excluding sessions with excessive electrical noise, 2,289 recording sessions and 729 decoding sessions were analyzed. T10’s MFG array was only used for recording evaluation, as it was in a different brain area with neural tuning properties not comparable to the hand knob area. High bandwidth neural data (30 kilo-samples s−1), required for retrospectively evaluating decoding performance in a uniform manner, were not consistently available for participants S1, S2, S3, A1 and T1, so these participants were not included in decoding analyses. Panel d adapted from ref. 62, © 2020, with permission from IOP Publishing.

To summarize findings spanning 14 participants over 20 years, here, we curated two types of datasets: (1) a ‘yield dataset’ containing a single neural recording from every research session, which was used to estimate array yield (Fig. 1f), and (2) a ‘decoding dataset’ consisting of all neural recordings during which a BCI was used to move a computer cursor toward well-defined targets (Fig. 1f), which was used to estimate neural decoding accuracy, neural tuning stability and electrode count scaling properties. This dataset consists of a more limited number of days and participants, because not every experimental session included BCI cursor control and because, for the first five participants, high-bandwidth neural data (30 kHz) was either not recorded during cursor control (S1, S2 and A1), not recorded for the first year post-implant (S3) or could not be consistently aligned to cursor task data (T1). Note that prior publications have shown successful demonstrations of neural cursor control in S1 and S223, S31,13,24,25,26, A124 and T125.

Neural spiking activity persists for multiple years with little average decline

We first investigated how the number of electrodes able to detect spiking activity (‘spiking electrodes’) changed over time, looking across a total of 2,289 recording sessions. Spiking activity was defined as the detection of at least two spiking events per second when using a threshold of −4.5 times the robust standard deviation of the voltage signal (Supplementary Methods 1.4). We observed substantial variability across 20 arrays in 14 different participants (Fig. 2a), with some arrays sharply declining in yield within the first year (S2, S3 and T2) or having consistently low yield (T3), some arrays increasing over time (T7 medial and T8 lateral) and many arrays staying relatively constant or gradually declining over time (for example, T5, T6 and T11). Yield trajectories in the first year were not always predictive of yields over longer periods (for example, T7’s medial array increased in yield after the first year). Yield curves were estimated using locally weighted scatter plot smoothing (LOWESS)27. Although the cause of S2’s and T2’s sudden declines in yield are not known, S3’s decline coincided with and may have been triggered by mechanical trauma to the pedestal that occurred outside of research study sessions. T3’s array, which had a consistently very low yield, was believed to have had its wire bundle unintentionally placed under tension during surgery, which could have resulted in the recording tips being largely epicortical rather than intracortical.

Fig. 2: Long-term trends in neural spiking activity recording performance.
Fig. 2: Long-term trends in neural spiking activity recording performance.

a, The array yield over the recording duration of twenty 96-channel microelectrode arrays. Each point represents the number of electrodes detecting spiking activity in a single neural recording of a one minute duration or longer (one recording per session). Spiking electrodes are defined as electrodes detecting a threshold crossing rate greater than or equal to 2 Hz when using a −4.5 times the robust standard deviation threshold (Supplementary Methods 2.1). LOWESS curves are shown in black. For participants with two arrays, data points in the lighter shade represent the more medial array. Implant date is indicated above each participant panel. Participant S2’s array experienced a steep increase in yield after repair of the percutaneous pedestal (day 190), which involved placing conductive epoxy on each bond site to restore connections. Months later, the array yield declined sharply again (day 323), and it is unknown whether this was due to failure of the previous repair or an unrelated event. Participant S3 experienced an accidental forceful contact of the pedestal against her headboard on the evening of day 181, with a decrease in yield that slowly, partially recovered. b, LOWESS curves for the mean spiking electrode count over all arrays (thick black line) and individual arrays (colors). c, Overall deltas in yield. Deltas were calculated by taking the mean number of spiking electrodes in sessions recorded during the last 3 months of recording and subtracting the mean number of spiking electrodes recorded during the first 3 months post-implant. d, Array proportions exceeding a certain yield over the first 3 years of recording. If data were not recorded from an array during a certain time period, that array was not included in the calculation for that time period.

Averaged over all days, participants and arrays, the mean array yield (the percentage of electrodes recording spiking activity) was 35.6%. Arrays experienced a modest 7 percentage point average decline in yield between the first 3 months post-implant and the last 3 months of recording (a 41% yield in the first 3 months to a 34% yield in the last 3 months) (Fig. 2c). However, there was large variability in yield across the full set of session and array level measurements (standard deviation of 21.4 percentage points). A minority of arrays in a subset of participants contributed substantially to this variability, including the three arrays that showed a sharp decrease in yield over the first year (S2, S3 and T2) (Fig. 2a). Overall, results were favorable compared with prior work in nonhuman primates, with over 50% of arrays maintaining at least 20% yield over the first 3 years (compared with the study by Sponheim et al.15, where less than 15% of arrays maintained a 20% yield) (Fig. 2d). This demonstrates the potential of long-term, reliable detection of neural spike waveforms with planar microelectrode arrays in humans.

We also examined electrode impedance and found that it increased sharply post-implant, followed by a consistent and gradual decline (Extended Data Fig. 1), consistent with prior reports21,28. Threshold-crossing waveform amplitude and background noise amplitude also declined over time (Extended Data Fig. 2a,b), and smaller waveform amplitudes were associated with lower impedance values (Extended Data Fig. 2c). As both the threshold-crossing waveform and background noise amplitudes declined at similar rates, the waveform signal-to-noise ratio remained relatively constant over time (Extended Data Fig. 2d). This overall decline in signal amplitude and impedance may be due to insulation degradation21,28, particularly around the edges of the planar array where low impedances were sometimes observed progressively over time (for example, T2, T3, T5 and T6) (Extended Data Fig. 3). Nevertheless, these changes did not appear to affect detection of spiking signals or decoding performance (see below).

Movement intention can be decoded reliably for years in most arrays

Next, we examined the extent to which cursor movement intention could reliably be decoded from intracortical recordings, how decoding varied across arrays and how it compared with the accuracy of able-bodied cursor control using a computer mouse. A total of 729 sessions across 14 arrays and nine participants were analyzed to evaluate movement intention decoding during closed-loop cursor-control tasks (Fig. 1f).

To evaluate the amount of intended movement information in the neural signals recorded by an array, we quantified how well an offline linear decoder could transform neural activity into a ‘movement intention’ unit vector pointing from the cursor to the target using neural features from a short time window at the beginning of each movement. The decoder output from each trial can be visualized as a dot in 2D space and colored according to the ground truth direction; the more accurate the decoder is, the more the dots should form a tight ring whose colors form an orderly gradient (Fig. 3a). An accurate decoder output represents movement intention vectors that point more accurately at the target with a consistent magnitude, and would thus enable higher quality cursor control, given that cursor-control algorithms often use linear mappings between neural features and cursor velocity1,3,4,24,29.

Fig. 3: Retrospective assessment of movement dSNR across time.
Fig. 3: Retrospective assessment of movement dSNR across time.

a, dSNR and example dSNR plots. dSNR is a vector-based, movement intention decoding metric that quantifies how well movement intention can be decoded from all electrodes of a given array during a target-directed cursor-control task. To estimate dSNR, retrospectively computed linear decoder output within a 300-ms window is decomposed into a signal component (pointing toward the target) and a noise component to calculate a dSNR. Each point in the dSNR plots denotes a trial, color-coded by target location. Spike band power neural features were retrospectively computed in a uniform manner for all participants from whom high-bandwidth neural recordings were available for analysis. A linear decoder was then trained in a uniform manner to decode these neural features, allowing performance to be meaningfully compared across participants and sessions despite variability in the online decoding algorithm used at the time of data collection. b, Example dSNR plots for able-bodied cursor control using a computer mouse. Cursor trajectories from an initial push toward the target were used to calculate dSNR. c, dSNR over time. Each point represents a dSNR value for one session. LOWESS curves are shown in black for each array. d, LOWESS curves for the mean dSNR over all arrays (thick black line) and individual arrays (colors). e, dSNR distributions. Arrays are displayed in implant order. Box plots summarize dSNR values across closed-loop cursor-control sessions for each array (T2, n = 42 sessions; T3, n = 8; T6, n = 124; T7 lateral, n = 35; T7 medial, n = 35; T8 lateral, n = 58; T8 medial, n = 58; T9 lateral, n = 117; T9 medial, n = 117; T10 PCG, n = 57; T5 lateral, n = 92; T5 medial, n = 92; T11 lateral, n = 196; and T11 medial, n = 196). The center line denotes the median, box bounds denote the 25th and 75th percentiles, and whiskers denote the minimum and maximum values within 1.5× the interquartile range. The dotted line indicates mean dSNR across nine able-bodied participants using a computer mouse. f, The overall deltas in dSNR. Deltas were calculated by taking the mean dSNR in sessions recorded during the last 3 months of recording and subtracting the mean dSNR recorded during the first 3 months post-implant.

To quantify the quality of the offline decoder output, the decoding signal-to-noise ratio (dSNR) metric30 decomposes each decoder output vector into a signal and a noise vector (Fig. 3a) and takes the ratio of their magnitudes (Supplementary Methods 2.4). dSNR is independent of the particular task parameters, decoding architecture and signal processing used in real-time during data collection, and thus enables meaningful post hoc comparisons across a diversity of datasets that would otherwise not be possible via task performance metrics alone. It can also be used to quantify target-directed able-bodied mouse movements (Fig. 3b), contextualizing the neural dSNR values. To compute dSNR for able-bodied mouse movements, we used the endpoint of the computer cursor after an initial ‘ballistic’ center-out cursor movement to define the ‘decoder output’ vector.

There was wide variability in dSNRs across participants and arrays (Fig. 3c), although, on average, the dSNR increased by 0.54 across the duration of recording (ranging from approximately 1 to 7.3 years) (Fig. 3f). Most arrays maintained dSNR values greater than 1, indicating a persistent ability to decode meaningful movement intention signals. A minority of arrays had low dSNR at the end of the first year post-implant and did not recover dSNR over the course of study participation (T2, T3 and T11 lateral). High-performing arrays with peak dSNR values greater than 4.5 (T5 lateral, T5 medial, T11 medial) approached able-bodied dSNR values (average computer mouse dSNR of 6.29) (Fig. 3e). This means that movement direction vectors could be linearly decoded from high-performing arrays with accuracy 71% as high as able-bodied mouse movement. However, typical arrays yielded about one-third the dSNR of able-bodied computer mouse movement. In prior work2,31, dSNR values of 5.5 and 5.6 (T5) enabled typing speeds of 30.8 and 36.1 correct characters per minute, while dSNR values between 1.9 and 3.3 (T6, T7 and T9) enabled typing speeds between 13.5 and 24.0 correct characters per minute (Extended Data Fig. 4).

The modest improvement in dSNR we observed over time in some arrays may reflect increased consistency in generating neural control signals (learning) and/or refinement of the movement strategies used to drive the cursor. For example, participant T6 shifted from wrist-based movements to index finger and thumb movements to better align motor imagery with the neural tuning patterns recorded on the array. Future studies are needed to separate the effects of learning from those of motor imagery to clarify why performance may improve over time.

Next, we assessed whether the variability we observed in overall dSNR levels between arrays could be explained by differences in array yield. We found that mean dSNR was correlated with mean array yield (Pearson r = 0.70, P = 5.32 × 10−3) (Fig. 4), but that array yield did not fully account for the large variability of decoding performance between arrays, in part due to information contained in ‘spike band power’ features (for example, participant T6). The results presented above used spike band power features to decode movement intention, as spike band power (power contained in the 250–5,000-Hz band) can capture low-amplitude spiking signals that might otherwise be missed by threshold crossing features. We evaluated how dSNR varied as a function of neural features used for decoding, and found that spike band power led to higher performance on average than threshold crossing rates, even at lower thresholds (Extended Data Fig. 5). It is not yet known why spike band power features were able to achieve a relatively high dSNR for participant T6 despite a low yield, but did not do so to the same extent for other low-yield arrays. Future work is required to disentangle potential sources of this variability, which could include variations in tissue response to the implant.

Fig. 4: Relationship between array yield and dSNR.
Fig. 4: Relationship between array yield and dSNR.

Across arrays, mean dSNR and mean yield were positively correlated (Pearson r(12) of 0.70, 95% CI 0.27–0.90; two-sided P = 5.32 × 10−3). When excluding the T6 array, which had low array yield but high information content in spike band power features, this relationship was stronger (Pearson r(11) of 0.84, 95% CI 0.54–0.95; two-sided P = 3.03 × 10−4). Small, color-coded points show individual session observations and larger circles show the mean yield and mean dSNR for each array. The solid line shows the ordinary least-squares fit across the 14 array means and the dashed line shows the corresponding fit after excluding the T6 array.

Overall, these outcomes show that long-term persistence of meaningful decoding performance is typical and can occur even at low array yields, although absolute dSNR levels can vary widely across participants and arrays. These same trends were also observed when using an angular error metric to quantify decoding performance (Extended Data Fig. 6).

Neural representations of movement intention remain stable on short time scales

Next, we investigated the day-to-day changes in how movement intention is represented in the recorded neural activity. Neural tuning stability was analyzed by using a linear encoding model to describe how spike band power features were tuned to a movement intention vector that points from the cursor to the target. After fitting, tuning stability was assessed using cosine similarity between model coefficients fit on different days (Supplementary Methods 2.6). Only arrays in the decoding dataset for which movement intention could be decoded reliably (median dSNR >1) were analyzed (three arrays excluded), yielding a total of 679 recording sessions and 11 arrays from seven participants.

Day-to-day tuning stability exhibited a variety of patterns (Fig. 5a), owing in part to potential changes in motor imagery across days and/or tasks, varying task instructions and variable intervals between consecutive cursor-control sessions. Nevertheless, many arrays showed clear regimes of long-lasting tuning stability. Average stability as a function of time elapsed (Fig. 5b) was generally high within the time span of 1 month (7 of 11 arrays, tuning similarity >0.6). This is encouraging, as short-term stability could enable decoders to track changes through small, iterative unsupervised adjustments (although some sporadic, larger shifts we observed may still require explicit recalibration)12,32,33,34.

Fig. 5: Stability of neural tuning to movement intention across days.
Fig. 5: Stability of neural tuning to movement intention across days.

a, Tuning stability heatmaps. A linear encoding model was used to describe neural activity as a function of a point-at-target vector that describes the user’s movement intention during a cursor-control task. The tuning similarity between model coefficients for each session pair is shown in each (i,j) entry. Blocks of blue color indicate time periods of high stability. b, Tuning stability as a function of day separation. Tuning remains stable on the time scale of a month (7 of 11 arrays, tuning similarity >0.6), which could enable decoders calibrated on prior days to be used on future days without requiring recalibration. Line colors correspond to the array names and heatmap outlines in a. The mean tuning similarity across arrays is denoted by the thick black line.

Electrode count scaling provides a pathway to higher performance

Finally, we examined how decoding performance scales with the number of electrodes. We computed average dSNR across all cursor-control sessions, for each array and participant, as a function of electrode count by randomly subsampling from all 96 electrodes (n = 500 resamplings per electrode count). Only arrays providing meaningful movement intention decoding (median dSNR >1) were evaluated. dSNR scaled with the number of electrodes, but did so sublinearly, up to 192 electrodes for participants with dual arrays and 96 electrodes for participants with a single array (Fig. 6). The sublinear increase in dSNR was well captured by simple monotonic models, including a logarithmic model (dSNR versus log2 electrodes) with R2 = 0.993 and P = 1.53 × 10−17 for single arrays and R2 = 0.989 and P = 8.86 × 10−20 for combined arrays (Supplementary Fig. 4). These results are supportive of the many ongoing efforts toward higher channel count devices35,36,37, although it appears that for arrays with lower dSNR values (for example, T8), many orders of magnitude more electrodes would be required to reach able-bodied computer mouse performance.

Fig. 6: dSNR scales approximately logarithmically with electrode count.
Fig. 6: dSNR scales approximately logarithmically with electrode count.

For each array, dSNR was assessed using a random subset of electrodes to investigate how dSNR varies as a function of electrode count (for each electrode count, dSNR was averaged over 500 random subsets). For individual arrays, electrode counts ranged from 6 to 96. For participants with dual arrays, combined electrode counts ranged from 6 to 192.

Leave a Reply

Your email address will not be published. Required fields are marked *